100
Chapter 4
Calibration
Calibration Types and Characteristics
Simplified full 3-port
calibration
•Open
•Short
•Load
•Thru
Same as full 3-port calibration
S11, S21, S31,
S12, S22,
[S32], S13,
[S23], S33
(Part of thru
measurement
can be
skipped)
•High-accuracy 3-port
measurement
*7*4
•Simpler procedure by skipping thru
measurement
Full 4-Port
Calibration
ECal module
(2-port
/
4-port)
Following 48 error terms:
•Directivity (Ed1,Ed2,Ed3,Ed4)
•Isolation
(Ex21,Ex31,Ex41,Ex12,Ex32,Ex4
2,Ex13,Ex23,Ex43,Ex14,Ex24,Ex
34)
•Source Match (Es1,Es2,Es3,Es4)
•Load Match (El1,El2,El3,El4)
•Transmission Tracking
(Et21,Et31,Et41,Et12,Et32,Et42,E
t13,Et23,Et43,Et14,Et24,Et34)
•Reflection Tracking
(Er1,Er2,Er3,Er4)
S11,S21,S31,S
41,S12,S22,S3
2,S42,S13,S23
,S33,S43,S14,
S24,S34,S44
(All
S-parameters
at 4 ports)
•Highly accurate 4-port
measurement
•
Quick calibration with low
chance of operator error
•OPEN
•SHORT
•LOAD
•THRU
•Highly accurate 4-port
measurement
Simplified full 4-port
calibration
•Open
•Short
•Load
•Thru
Same as full 4-port calibration
S11, S21, S31,
[S41], S12,
S22, [S32],
[S42], S13,
[S23], S33,
S43, [S14],
[S24], S34,
S44
(Part of thru
measurement
can be
skipped)
•High-accuracy 4-port
measurement
•Simpler procedure by skipping thru
measurement
2-Port TRL
Calibration
•Reflection
(OPEN or
SHORT)
•THRU
•LINE
•MATCH
Following 12 error terms:
•Directivity (Ed1,Ed2)
•Source Match (Es1,Es2)
•Load Match (El1,El2)
•Transmission Tracking
(Et21,Et12)
•Reflection Tracking (Er1,Er2)
S11,S21,S12,S
22
(All
S-parameters
at 2 ports)
•Highly accurate 2-port
measurement
• Effective for non-coaxial device
measurement
3-Port TRL
Calibration
•Reflection
(OPEN or
SHORT)
•THRU
•LINE
•MATCH
Following 27 error terms:
•Directivity (Ed1,Ed2,Ed3)
•Source Match (Es1,Es2,Es3)
•Load Match
(El21,El31,El12,El32,El13,El23)
•Transmission Tracking
(Et21,Et31,Et12,Et32,Et13,Et23)
•Reflection Tracking (Er1,Er2,Er3)
S11,S21,S31,S
12,S22,S32,S1
3,S23,S33
(All
S-parameters
at 3 ports)
•Highly accurate 3-port
measurement
•Effective for non-coaxial device
measurement
Table 4-1
Calibration Types and Characteristics
Calibration Method
Standard(s)
Used
Corrected Error Factor
Measurement
Parameters
Characteristics
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...