32
Chapter 2
Troubleshooting
Troubleshooting Using Diagnostic Test
Step 6.
Double-click "Service" folder to open it.
Step 7.
Click "DiagnosticTest. VBA" program file to select it from program menu.
Step 8.
Click "Open" button to download the diagnostic test program.
Step 9.
Press
Select Macro
to select
Select Macro
function.
Step 10.
Press
Module1 main
to open the
Module1 main
program file. "Diagnostic Test Program"
dialog box (Main Menu) will appear as shown in Figure 2-14.
Figure 2-14
Diagnostic Test dialog box
Step 11.
To exit the diagnostic test, click "Exit" button.
Program Overall
Instrument
Information
The option and serial number for the E5061B, and test date can be
entered automatically.
Test Conditions
The test conditions, “Temperature”, “Humidity” and person doing the
test (“Tested by”) can be entered.
Test Items
The following test can be selected. To execute the test you select, click
“Execute” button.
•
DC-BUS Test
•
Internal Level Monitor Test
•
RF Output Level Range Test
•
RF Output Level Power SweepTest
•
Receiver Absolute Measurement Test
•
Receiver Compression Test (Option 3L5 only)
•
Receiver IF Ranging Test
•
Gain-Phase Absolute Measurement Test (option 3L5 only)
When each test is executed, the test dialog box is displayed. The
following screen is an example of the test dialog box.
Summary of Contents for E5061B
Page 1: ... Established 1981 Advanced Test Equipment Rentals www atecorp com 800 404 ATEC 2832 ...
Page 51: ...50 Chapter3 Replaceable Parts Replaceable Parts List ...
Page 103: ...102 Chapter4 Replacement Procedure Handle Assembly Replacement ...
Page 116: ...115 E Power Requirement D Power Requirement ...
Page 123: ...122 AppendixE Messages ...