104
Chapter 5
Post-Repair Procedures
Post-Repair Procedures
Post-Repair Procedures
Post Repair Procedures
lists the required procedures that must be performed
after the replacement of an assembly. These are the recommended minimum procedures to
ensure that the replacement is successfully completed.
Table 5-1
Post-Repair Procedures (Option 3L5)
Replaced
Assembly or
Part
Required Adjustments
Correction Constants (CC)
Verification
A1 Source
Module
Perform the following required
adjustments using “A1 Source Module” in
Spot Adjustment of the program.
“To Execute the Diagnostic Test” on page 31
Frequency Accuracy Test
RF Output Level Accuracy and Flatness Test
RF Output Level Linearity Test
Trace Noise Test
Crosstalk & System Dynamic Range Test
Dynamic Accuracy Test
Uncorrected System Performance Test
LF Output Level Accuracy Test (Opt 3L5)
LF Output Level Linearity Test (Opt 3L5)
Input Impedance Test (Opt 3L5)
Absolute Amplitude Accuaracy Test (Opt
3L5)
Magnitude Ratio/Phase Frequency Response
Test (Opt 3L5)
Trace Noise Test (Opt 3L5)
Crosstalk and Noise Level Test (Opt 3L5)
Magnitude Ratio/Phase Dynamic Accuracy
Test (Opt 3L5)
Summary of Contents for E5061B
Page 1: ... Established 1981 Advanced Test Equipment Rentals www atecorp com 800 404 ATEC 2832 ...
Page 51: ...50 Chapter3 Replaceable Parts Replaceable Parts List ...
Page 103: ...102 Chapter4 Replacement Procedure Handle Assembly Replacement ...
Page 116: ...115 E Power Requirement D Power Requirement ...
Page 123: ...122 AppendixE Messages ...