24
Chapter 2
Upgrade Procedure
Step 5: Uninstalling the E5022A/B System Software from your PC
Step 5: Uninstalling the E5022A/B System Software from
your PC
You had better to uninstall the old system software before installing a new system software.
This is because the structure of the start menu and some register information have been
changed and the uninstaller of the new system software might not delete them.
1. Select
Settings > Control Panel
from the windows start menu.
2. Select
Add/Remove Programs
.
3. Select
Agilent E5022/E5023
, then click
Add/Remove
to uninstall the system software.
Step 6: Installing the E5022/23 System Software into your PC
Install the appropriate E5022/23 system software into your PC.
Step 7: Installing the Compensation Factor
Execute the program named “caltool.exe” in order to install the calibration loss factor into
the PC. You can download this software from the website of “Service” at the URL
“http:\\dst.tm.agilent.com\”.
1. Execute the “caltool.exe” program on the PC.
2. The Figure 2-8 is displayed. The current stored value is displayed in the text box.
Figure 2-8
Caltool Dialogue
3. Input the value for the cable compensation in the text box, then click the
Store
button.
The value of the provided cables for E5023A is
-
0.00004041
.
For Upgrading To
Required System Software Revision
E5023A Option 415 (1.5 Gbps)
B.01.00 and above
E5023A Option 426 (2.6 Gbps)
B.03.00 and above
Summary of Contents for E5023U