32
Chapter 1
Overview
Service Equipment You Will Need
Step Attenuator Loss Characterization
The step attenuator combination should have each attenuator setting
characterized by a metrology lab at 50 MHz. The following tables show
which sections of the 10 dB and 1 dB step attenuators are utilized for
each attenuator setting. The tables also list the Allowable Uncertainty
for each attenuator setting. The interconnect cable should NEVER be
disconnected once the loss characterization is performed.
b. Note: One PSG with Option 567, 1EA, 1E1, 007, and UNX or UNR can be used as the Microwave Source #1, RF
Source #2, and RF Source #3.
(Option 007, Analog Ramp Sweep, required for Frequency Response Adjustments)
c. Supported PSG models:
E8244A
E8254A
E8257C Option H31 or 540
E8257D
E8267C Option H31 or 544
E8267D
(One sweeper requires Option 1E1 for PSA Option 110 testing)
d. PSG requires Option UNX or UNR.
e. The 8482A power sensor uses cal factors to compensate the power sensor for frequency response errors. Cal fac-
tors are stated in percentages. The 8482A factory cal factor uncertainty ranges from 2.2% to 3.1%. The cal factor
uncertainty can be reduced to < 2.0% by using metrology grade calibration techniques. The power sensor cal fac-
tor uncertainty becomes one component of the Verification Test uncertainty analysis. Lower cal factor uncer-
tainties will translate to wider test margins.
f.
The step attenuators should be permanently joined via the 11716A Interconnect Kit as shown in the diagram.
Summary of Contents for E4440A
Page 12: ...12 Contents...
Page 13: ...13 1 Overview...
Page 42: ...42 Chapter1 Overview Contacting Agilent Technologies...
Page 43: ...43 2 Overall Troubleshooting...
Page 69: ...69 3 Troubleshooting the RF Section E4440A E4443A E4445A...
Page 89: ...89 4 Troubleshooting the RF Section E4446A E4447A E4448A...
Page 111: ...111 5 Troubleshooting the Synthesizer Section...
Page 125: ...125 6 Troubleshooting the IF Section...
Page 144: ...144 Chapter6 Troubleshooting the IF Section Verifying the A7 Digital IF Board...
Page 145: ...145 7 Troubleshooting the Processor Power Supply and Display...
Page 165: ...165 8 Hardware Options...
Page 184: ...184 Chapter8 Hardware Options Verifying Option 122 or 140 Wide Bandwidth Digitizer...
Page 186: ...186 Chapter8 Hardware Options Verifying Option 122 or 140 Wide Bandwidth Digitizer...
Page 206: ...206 Chapter8 Hardware Options Verifying Option 124 Y Axis Video Out...
Page 207: ...207 9 Block Diagrams...
Page 211: ...Chapter 9 211 Block Diagrams Signal Mnemonics Figure 9 1 Graphic Symbols...
Page 212: ...212 Chapter9 Block Diagrams Overall Block Diagrams Overall Block Diagrams...
Page 214: ...214 Chapter9 Block Diagrams Overall Block Diagrams...
Page 216: ...216 Chapter9 Block Diagrams Overall Block Diagrams...
Page 217: ...Agi l ent MOTHERBOARD E4440 60001 se860a...
Page 218: ...218 Chapter9 Block Diagrams Overall Block Diagrams...
Page 220: ...220 Chapter9 Block Diagrams Overall Block Diagrams...
Page 221: ...221 10 Replaceable Parts Lists and Locations...
Page 252: ...252 Chapter10 Replaceable Parts Lists and Locations Hardware Figure 10 4 Major Assemblies...
Page 256: ...256 Chapter10 Replaceable Parts Lists and Locations Hardware Figure 10 6 YTO Assembly...
Page 267: ...Chapter 10 267 Replaceable Parts Lists and Locations Hardware...
Page 292: ...292 Chapter10 Replaceable Parts Lists and Locations Hardware Figure 10 27 Front Panel Parts...
Page 298: ...298 Chapter10 Replaceable Parts Lists and Locations Hardware Figure 10 32 Rear Frame Hardware...
Page 304: ...304 Chapter10 Replaceable Parts Lists and Locations Hardware...
Page 305: ...305 11 Assembly Replacement Procedures...
Page 383: ...Chapter 11 383 Assembly Replacement Procedures A25 Motherboard Figure 11 47 Bottom Screws...
Page 409: ...409 12 Post Repair Procedures...
Page 429: ...429 13 Firmware Upgrades...
Page 432: ...432 Chapter13 Firmware Upgrades Firmware Upgrades Including Measurement Personality Upgrades...