Chapter 2
29
E4438C Vector Signal Generator Overview
Front Panel Overview
13. LF OUTPUT Connector
This BNC connector is the output for modulation signals generated by the low frequency (LF) source
function generator. This output is capable of driving 3 V
p
(nominal) into a 50
Ω
load.
On signal generators with Option 1EM, this output relocated to a rear-panel SMB connector.
14. RF OUTPUT Connector
This female Type-N connector is the output for RF signals. The source impedance is 50
Ω
. For Options 501,
502, 503, and 504 the damage levels are 50 Vdc, 50 W at
≤
2 GHz, and 25 W at > 2 GHz maximum. For
Options 501, 502, 503, and 504 the reverse power protection circuit will trip, however, at nominally 1 W.
CAUTION
E4428C and E4438C signal generators with Option 506 are not equipped with reverse
power protection circuits.
On signal generators with Option 1EM, this output relocated to a rear-panel female Type-N connector.
15. Mod On/Off Key
Pressing this hardkey enables or disables all active modulation formats (AM, FM,
Φ
M, Pulse, or I/Q) that
are applied to the output carrier signal.
This hardkey does not set up or activate an AM, FM,
Φ
M, Pulse, or I/Q format; each individual modulation
format must still be set up and activated (for example,
AM > AM On
) or nothing will be applied to the output
carrier signal when the
Mod On/Off
hardkey is enabled.
The
MOD ON/OFF
annunciator, which is always present on the display, indicates whether active modulation
formats have been enabled or disabled with the
Mod On/Off
hardkey.
16. RF On/Off Key
This hardkey toggles the operating state of the RF signal present at the RF OUTPUT connector. The
RF On/Off
annunciator is always visible in the display to indicate whether the RF is turned on or off.
17. Numeric Keypad
The numeric keypad consists of the 0 through 9 hardkeys, a decimal point hardkey, and a backspace hardkey
(
). The backspace hardkey enables you to backspace or specify a negative value. When specifying a
negative numeric value, the negative sign must be entered prior to entering the numeric value.
Summary of Contents for E4428C
Page 22: ...Contents xxii ...
Page 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Page 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Page 229: ...205 6 Analog Modulation ...
Page 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Page 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Page 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Page 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Page 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Page 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Page 667: ...643 18 Troubleshooting ...