10
Chapter 2
Getting Started
Meeting Electrical and Environmental Requirements
Line Settings
The signal generator has an autoranging line voltage input. The available ac power source must meet the
following conditions:
Voltage
:
100/115 volts nominal (90-132 volts)
230/240 volts nominal (198-254 volts)
Frequency
: for 100/115 volts: 50/60 Hz nominal
50/60/400 Hz nominal (instruments with serial prefixes
>US4907/MY4907)
for 230/240 volts: 50/60 Hz nominal
50/60/400 Hz nominal (instruments with serial prefixes
>US4907/MY4907)
Power
:
200 watts maximum
CAUTION
Damage to the signal generator may result if the supply voltage is not within the specified
range.
Connecting the AC Power Cord
This is a Safety Class 1 Product provided with a protective earth ground incorporated into the power cord.
The front panel switch is only a standby switch; it is not a line switch. The ac power cord is the
disconnecting device that disconnects the signal generator mains circuits from the mains supply.
Alternatively, an external switch or circuit breaker, readily identifiable and easily reached by the operator,
may also be used as a disconnecting device.
Perform the following steps when connecting the ac power cord:
WARNING
Personal injury may occur if there is any interruption of the protective conductor
inside or outside of the signal generator. Intentional interruption is prohibited.
CAUTION
Damage to the signal generator may result without adequate earth grounding. Always use
the three-prong ac power cord supplied with the signal generator.
Summary of Contents for E4428C ESG
Page 7: ...1 1 Safety Information ...
Page 12: ...6 Chapter 1 Safety Information General Safety Considerations ...
Page 13: ...7 2 Getting Started ...
Page 28: ...22 Chapter 2 Getting Started Contacting Agilent Sales and Service Offices ...
Page 29: ...23 3 Operation Verification ...
Page 61: ...55 4 Regulatory Information ...
Page 64: ...58 Chapter 4 Regulatory Information Regulatory Information misc ...