Power Meter Operation
36
E4418B Power Meter User’s Guide
Effective control of the zero and calibration cycles using the TTL inputs
depends on correct timing of the input signals as shown in
and
Table 2-3
TTL Input Timing Diagram 1
Timing of zero/cal inputs for conditions “01” and “10”.
Time
Description
Value
T1
Minimum width of input
300 ms
T2
Time between input detection and start of zero/cal
cycle. This is determined by the number of averages
x sample rate or if there is an existing zero/cal
operation in progress, the time for this current
operation to complete. Note that the worst case is
1024 averages x 50 ms = 51.2 s. For front panel
operation (in free run mode) the time is 1 x 50 ms.
Max: 50 ms (typical)
Min: 0 ms
T3
Maximum width of input. Longer inputs may cause a
subsequent zero/cal operation some time after the
current one is complete.
4 s
T1
T2
T3
T4
Input
Operation
Summary of Contents for E4418B
Page 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Page 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...