![Agilent Technologies E4418B Service Manual Download Page 56](http://html.mh-extra.com/html/agilent-technologies/e4418b/e4418b_service-manual_2868036056.webp)
Performance Tests
Agilent E-Series Power Sensor Functional Test
2-20
Agilent E4418B/E4419B Service Guide
Agilent E-Series Power Sensor Functional Test
Description
This test verifies that the meter/sensor combination can make RF
measurements.
The Agilent E-series power sensors operate over a 90 dB dynamic range
(-70 to +20 dBm). However, since there is an amplifier in these power
sensors, the voltages presented to the power sensor connector are always
within the range of voltages available from the Agilent 11683A range
calibrator.
Equipment
Power Sensor Cable............................... Agilent 11730A
Reference Attenuator ............................ Agilent 11708A
Agilent E-series power sensor
Procedure
To complete the Functional Test the following procedure should be
performed.
Note
The procedure details the key presses required on the
Agilent E4418B. For the Agilent E4419B the equivalent key
presses should be performed on both channels.
1.
Connect the equipment as shown in Figure 2-5.
Figure 2-5: Sensor Functional Test Setup
Power Meter
Agilent E-Series
Power Sensor
POWER
REF
CHANNEL A
Agilent 11730A
4402serv.book Page 20 Tuesday, October 14, 2003 3:18 PM
Summary of Contents for E4418B
Page 19: ...1 Specifications 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 37: ...2 Performance Tests 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 69: ...3 Adjustments 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 81: ...4 Theory of Operation 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 95: ...5 Replaceable Parts 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 119: ...6 Troubleshooting 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...