Performance Tests
Introduction
Agilent E4418B/E4419B Service Guide
2-3
Equipment Required
Table 2-1 lists all the equipment required for the adjustments and
performance tests. If substitutions must be made, the equipment used
must meet the critical specifications.
Table 2-1: Required Equipment
Equipment
Required
Model Number
Critical Specification
Usage
1
1. P = Performance Tests, A = Adjustments
Digital Multimeter
Agilent 3458A
Range 0 to 20 V
Resolution 0.01 V
P and A
DC Calibrator
Datron 4000A
or
Fluke 5720A
P
Range Calibrator
Agilent 11683A
or
Agilent 11683A
Opt H01
Option H01 is essential for
accurate verification. (See
analysis in
). The
Standard 11683A can be used
for indication only.
P
Frequency Counter
Agilent 53132A
Range 10 Hz to 50 MHz
Resolution 1 Hz
P and A
Power Meter
Agilent 432A
Range 1 mW
Transfer Accuracy 0.2%
P and A
Thermistor Mount
Agilent 478A Opt H75/H76
or
Agilent 8478B Opt H01
SWR<1.05 at 50 MHz
Accuracy
±
0.2% at 50 MHz.
P and A
Power Sensor
Agilent E-series power
sensor
P
30 dB Attenuator
Agilent 11708A
or
Agilent 8491A Option 030
P
Power Sensor
Cable
Agilent 11730A
or
Agilent
E9288A
P and A
4402serv.book Page 3 Tuesday, October 14, 2003 3:18 PM
Summary of Contents for E4418B
Page 19: ...1 Specifications 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 37: ...2 Performance Tests 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 69: ...3 Adjustments 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 81: ...4 Theory of Operation 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 95: ...5 Replaceable Parts 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...
Page 119: ...6 Troubleshooting 4402serv book Page 1 Tuesday October 14 2003 3 18 PM ...