11. Verify that the test spectrum analyzer signal trace is
≥
5 dBm
between 2.1 GHz and 3.1 GHz. Refer to Figure 1.
Figure 1-4
Option HSB Trace Characteristics
12. Press
[Span]
>
Zero Span
on the ESA DUT.
13. Press
[Center Frequency] [0 MHz]
on the ESA DUT.
14. Press
[View/Trace]
>
Clear Write
on the test spectrum analyzer.
15. Set the
[Marter]
on the test spectrum analyzer to 1.8214 GHz.
16. Verify that the test spectrum analyzer signal trace is < –15 dBm at
1.8214 GHz. The HSB board LO signal must be attenuated
sufficiently for proper sideband adapter operation.