Appendix Service Information
Recommended Test Equipment
179
Service
Information
Recommended Test Equipment
The test equipment recommended for the performance verification and adjustment
procedures is listed below. If the exact instrument is not available, use the accuracy
requirements shown to select substitute calibration standards. If you use equipment
other than that recommended in Table A-2, you must recalculate the measurement
uncertainties for the actual equipment used.
Table A-2 Recommended Test Equipment
*
To find the accurate resistance, it is recommended to use a current monitoring resistor after calibration.
1
E3646A model,
2
E3647A model,
3
E3648A model,
4
E3649A model.
Instrument
Requirements
Recommended Model
Use
GPIB controller
Full GPIB or RS-232
capabilities
Agilent 82341C interface
card or equivalent
Programming and readback
accuracy
Oscilloscope
100 MHz with 20 MHz
bandwidth
Agilent 54602B
Display transient response and
ripple & noise waveform
RMS Voltmeter
20 Hz to 20 MHz
Measure rms ripple & noise
Cable (BNC to BNC)
50 ohm, 9 inch (23 Cm)
Agilent 10502A or 10503A if
the 10502A is not available.
Measure rms ripple & noise
(CV PARD, CC PARD)
BNC (Female)
Bulkhead Receptacle
Isolated Ground.
Nominal impedance: 50 Ohm
Pomona Model 5148
Measure rms ripple & noise
(CV PARD, CC PARD)
Split Ferrites
For use with round cable
Steward Co. 28A2029-0A0 Noise coupling reduction
Digital Voltmeter
Resolution: 0.1 mV
Accuracy: 0.01%
Agilent 34401A
Measure dc voltages
Electronic Load
Voltage Range: 240 Vdc
Current Range: 10 Adc
Open and Short Switches
Transient On/Off
Agilent 60503B
Measure load and line
regulations and transient
response time.
Resistive Loads (R
L
)
(2.7
Ω
, 150 W/13.5
Ω
, 150 W)
1
(43.8
Ω
, 300 W/120
Ω
, 300 W)
2
(1.6
Ω
, 300 W/8.0
Ω
, 300 W)
3
(25
Ω
, 300 W/75
Ω
, 300 W)
4
Measure ripple and noise
Current monitoring
Resistor (Shunt) - R
M1
(0.01
Ω
, 0.1%)
*
ISOTEK Co. Model: A-H or
equivalent
Constant current test setup
Current monitoring
Resistor (Shunt) - R
M2
(0.1
Ω/ 5
W
, 1%)
*
Vishay Co. Model: LVR-5-
43 or equivalent
Measure current rms ripple &
noise
Summary of Contents for E3646A
Page 2: ...2...
Page 16: ...16 Contents Contents...
Page 17: ...1 Quick Start...
Page 27: ...2 General Information...
Page 39: ...3 Front Panel Operation and Features...
Page 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Page 73: ...4 Remote Interface Reference...
Page 125: ...5 Error Messages...
Page 136: ...Chapter 5 Error Messages Calibration Errors 136...
Page 137: ...6 Application Programs...
Page 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Page 149: ...7 Tutorial...
Page 160: ...Chapter 7 Tutorial Remote Programming 160...
Page 161: ...8 Specifications...
Page 169: ...Appendix Service Information...
Page 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Page 210: ......