Advanced Software Features
15
Agilent E5505A User’s Guide
347
PLL Suppression Verification Process
When “Verify calculated phase locked loop suppression” is selected, it is
recommended that “Always Show Suppression Graph” also be selected.
Verifying phase locked loop suppression is a function which is very useful in
detecting errors in the phase detector constant or tune constant, the tune
constant linearity, limited VCO tune port bandwidth conditions, and injection
locking conditions. If the DUT is well behaved (injection locking issues do not
exist or have been eliminated) and the reference source is well behaved (well
known tuning characteristics or a system controlled RF signal generator) then
the need to select PLL suppression verification is minimal.
To verify PLL suppression, a stimulus source is required for the FFT analyzer.
This stimulus signal is connected to the CHIRP INPUT port on the rear panel
of the test set. The PC digitizer, when used as the FFT analyzer, provides a
companion D/A output for this purpose. When an Agilent 89410A vector signal
analyzer is the system FFT analyzer, the 89410A’s companion source output is
used.
PLL suppression information
The PLL Suppression View graph has been updated to allow measured,
calculated (adjusted), and theoretical information to be examined more
closely. When the “Always Show Suppression Graph” is selected, the following
graph (
) is provided.
Figure 273
Default PLL suppression verification graph
e5505a_user_select_suppression.ai
rev2 10/21/03
Summary of Contents for E1420B
Page 24: ...24 Agilent E5505A User s Guide ...
Page 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Page 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Page 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Page 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Page 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Page 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Page 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Page 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Page 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Page 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Page 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...