Hub Upstream Signal Quality Testing
4
DSOX4USBSQ USB 2.0 Signal Quality Analysis Application Electrical Testing Notes
59
Hub Upstream Hi-Speed Compliance Test Packets
To set up the Host computer to transmit compliance test packets, perform
these steps:
1
Invoke the HS Electrical Test Tool software on the Hi- Speed Electrical
Test Bed computer. Select
Hub
and click the
TEST
button to enter the
Hub Test dialog box.
2
In the Hub Test dialog box, the hub under test should be enumerated
with the hub's VID shown together with the USB address.
Select
TEST_PACKET
from the
Hub Command
drop down menu.
3
Click
EXECUTE
.