Page 9 of 9
MSO6104A-01A-S
Ensure all hooks and tabs are inserted correctly in their slots.
Figure 7
Insert hook legs and lock.
To reinstall the cabinet
1.
Slide the oscilloscope into the cabinet until connectors are fully in their openings in the rear
panel.
2.
Replace the two T20 screws shown in Figure 1 on page 3.
Run user self test
Follow the instructions in the service guide to verify that the instrument passes User Self Test. No
calibration is required.
Insert Lock