Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
365
:TRIGger[:EDGE]:COUPling
(see
Command Syntax
:TRIGger[:EDGE]:COUPling <coupling>
<coupling> ::= {AC | DC | LFReject}
The :TRIGger[:EDGE]:COUPling command sets the input coupling for the
selected trigger sources. The coupling can be set to AC, DC, or LFReject.
•
AC coupling places a high- pass filter (10 Hz for analog channels, and
3.5 Hz for all External trigger inputs) in the trigger path, removing dc
offset voltage from the trigger waveform. Use AC coupling to get a
stable edge trigger when your waveform has a large dc offset.
•
LFReject coupling places a 50 KHz high- pass filter in the trigger path.
•
DC coupling allows dc and ac signals into the trigger path.
Query Syntax
:TRIGger[:EDGE]:COUPling?
The :TRIGger[:EDGE]:COUPling? query returns the current coupling
selection.
Return Format
<coupling><NL>
<coupling> ::= {AC | DC | LFR}
See Also
•
"Introduction to :TRIGger Commands"
•
•
N O T E
The :TRIGger[:EDGE]:COUPling and the :TRIGger[:EDGE]:REJect selections are coupled.
Changing the setting of the :TRIGger[:EDGE]:REJect can change the COUPling setting.
Summary of Contents for DSO6014A/L
Page 1: ...A Agilent 6000 Series Oscilloscopes Programmer s Reference...
Page 16: ...16 Agilent 6000 Series Oscilloscopes Programmer s Reference...
Page 30: ...30 Agilent 6000 Series Oscilloscopes Programmer s Reference 1 What s New...
Page 540: ...540 Agilent 6000 Series Oscilloscopes Programmer s Reference 6 Error Messages...
Page 562: ...562 Agilent 6000 Series Oscilloscopes Programmer s Reference 7 Status Reporting...
Page 624: ...624 Agilent 6000 Series Oscilloscopes Programmer s Reference 9 Programming Examples...