356
Agilent 6000 Series Oscilloscopes Programmer's Reference
3
Commands by Subsystem
:TRIGger:DURation:LESSthan
(see
Command Syntax
:TRIGger:DURation:LESSthan <less than time>[<suffix>]
<less than time> ::= maximum trigger duration in seconds
(5 ns - 10 seconds) in NR3 format
<suffix> ::= {s | ms | us | ns | ps}
The :TRIGger:DURation:LESSthan command sets the maximum duration for
the defined pattern when :TRIGger:DURation:QUALifier is set to LESSthan.
Query Syntax
:TRIGger:DURation:LESSthan?
The :TRIGger:DURation:LESSthan? query returns the duration time for the
defined pattern.
Return Format
<less than time><NL>
See Also
•
"Introduction to :TRIGger Commands"
•
•
•
Summary of Contents for DSO6014A/L
Page 1: ...A Agilent 6000 Series Oscilloscopes Programmer s Reference...
Page 16: ...16 Agilent 6000 Series Oscilloscopes Programmer s Reference...
Page 30: ...30 Agilent 6000 Series Oscilloscopes Programmer s Reference 1 What s New...
Page 540: ...540 Agilent 6000 Series Oscilloscopes Programmer s Reference 6 Error Messages...
Page 562: ...562 Agilent 6000 Series Oscilloscopes Programmer s Reference 7 Status Reporting...
Page 624: ...624 Agilent 6000 Series Oscilloscopes Programmer s Reference 9 Programming Examples...