Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
211
:EXTernal:UNITs
(see
Command Syntax
:EXTernal:UNITs <units>
<units> ::= {VOLTs | AMPeres}
The :EXTernal:UNITs command sets the measurement units for the probe
connected to the external trigger input. Select VOLTs for a voltage probe
and select AMPeres for a current probe. Measurement results, channel
sensitivity, and trigger level will reflect the measurement units you select.
Query Syntax
:EXTernal:UNITs?
The :CHANnel<n>:UNITs? query returns the current units setting for the
external trigger.
Return Format
<units><NL>
<units> ::= {VOLT | AMP}
See Also
•
"Introduction to :EXTernal Trigger Commands"
•
"Introduction to :TRIGger Commands"
•
•
•