Agilent Technologies DNA 1000 Kit Quick Start Manual Download Page 3

3

Agilent DNA 1000 Kit Quick Start Guide

Setting up the Chip Priming Station

1

Replace the syringe:

a

Unscrew the old syringe from the lid of the chip priming station.

b

Release the old syringe from the clip. Discard the old syringe.

c

Remove the plastic cap of the new syringe and insert it into the clip.

d

Slide it into the hole of the luer lock adapter and screw it tightly to the chip 
priming station.

2

Adjust the base plate:

a

Open the chip priming station by pulling the latch.

b

Using a screwdriver, open the screw at the underside of the base plate.

c

Lift the base plate and insert it again in position C. Retighten the screw.

3

Adjust the syringe clip:

a

Release the lever of the clip and slide it down to the lowest position.

Essential Measurement Practices

Handle and store all reagents according to the instructions on the label of the 
individual box.

Avoid sources of dust or other contaminants. Foreign matter in reagents and 
samples or in the wells of the chip will interfere with assay results.

Keep all reagents and reagent mixes refrigerated at 4 °C when not in use.

Allow all reagents and samples to equilibrate to room temperature for 30 min before use.

Protect dye and dye mixtures from light. Remove light covers only when pipetting. The dye decomposes 
when exposed to light and this reduces the signal intensity.

Always insert the pipette tip to the bottom of the well when dispensing the liquid. Placing the pipette at 
the edge of the well may lead to poor results.

Use a new syringe and electrode cleaners with each new kit.

Use loaded chips within 5 min after preparation. Reagents might evaporate, leading 
to poor results. 

Do not touch the Agilent 2100 Bioanalyzer during analysis and never place it on a 
vibrating surface.

Agilent DNA Assay Protocol 

WA R N I N G

Handling DMSO
Kit components contain DMSO. Because the dye binds to nucleic acids, it should be treated as a 
potential mutagen and used with appropriate care. 

Wear hand and eye protection and follow good laboratory practices when preparing and handling 
reagents and samples.

Handle solutions with particular caution as DMSO is known to facilitate the entry of organic 
molecules into tissues. 

Reviews: