39
Setting up the computer and software
Verifying the profile settings and configuration
AssayMAP Bravo Platform Installation Guide
Verifying the profile settings and configuration
Ensure that you complete the following workflow for all device profiles in the
Protein Sample Prep Workbench.
After you ensure that all profiles in the Protein Sample Prep Workbench have
been installed and that they are configured correctly, you set the teachpoints
for each of the profiles.
Verifying the profile settings
IMPORTANT
Ensure that the Protein Sample Prep Workbench is closed.
To verify the profile settings:
1
Start the VWorks software.
You can click the VWorks desktop icon or use the Start menu on the
Windows desktop to start the VWorks software.
2
Ensure that Simulation is off.
Figure
Simulation button in the VWorks window
3
In the
VWorks
window, open the corresponding AssayMAP device file for
the profile that you are verifying.
To open a device file, choose
File > Open
. In the
Open
dialog box, ensure the
file type is
Device Files (*.dev)
, select the file, and then click
Open
.
The device files are installed in the following default location:
C:\VWorks Workspace\Workbench\Device Files
4
Click
No
when the
Would you like to initialize devices now?
message appears.
Step
For this task...
See...
1
Verify the device file
references the correct profile
and the profile settings are
correct.
“Verifying the profile settings” on
page 39
2
Ensure the deck configuration
in the profile matches the
physical AssayMAP Bravo
deck.
“About verifying the deck
configuration in the profile” on
page 43