background image

3) 

Figure 2

 

Opened shipping container with the 

accessory

 

tray

 sitting on top

7) Remove the plastic wrapping from the instrument.
8) Lift the instrument off of the shipping container base and set it down on its selected location

(see “Step 1. Select a location for the instrument”). Agilent recommends that two people lift the instrument together.

1) 

Figure 4 

Foam insert (

left

) and underlying cardboard strip (

right

)

1) 

2) 

Unpack the shipping containers 

step

2

Figure 1

 

Aria

D

x shipping container

Turn over to continue

4)  
5) 

6) 

Locate a solid, flat clean surface for the instrument. Make sure that:
• The instrument can stand completely stable.
• The rear air slots will not be covered.
•  The unit always has at least 10 cm (approximately 4 inches) to the next wall or neighboring instrument.
•  The temperature (normal ambient) is between 20°C and 30°C with humidity levels between 20% and 80% non-condensing.

 

• The atmosphere is not explosive.

Select a location for the instrument 

step

1

Install the optical modules 

step

3

Figure 3

 

Instrument on shipping container base, with box

sleeve removed

Installing the Aria

D

x Instrument

2) 

 Open the instrument door by lifting up on the handle on the top of the instrument. Lift the door all the way up and back. 

Remove the piece of foam and then remove the strip of cardboard from around the thermal block assembly (see Figure 4).

 

The optical module housing carrier is positioned to the left of the thermal block assembly. 

Slide the optical module housing carrier to the right until it is centered in the opening of the instrument door (as shown in 

Figure 5). Use the indentation on the top of the carrier to help slide it.

3) 

The Aria

D

x instrument is shipped in two separate containers.  

The small container holds the power cord. The large container 

holds the instrument and accessory tray.

Optical modules are packed separately.

Open the small shipping container. Remove the power cord 

and set it aside for now.

Make sure that the large shipping container is in the upright 

position (Figure 1), then cut the four plastic straps that hold 

the container together.

Open the top flaps of the large container. Inside the container is an accessory tray (Figure 2), which contains the Certificate of 

Conformance and the installation poster.

Remove the accessory tray and set it aside.

Remove the foam support that sits on top of the instrument in the shipping container.

Grip a handle on each side of the shipping container and lift up to remove the box sleeve that surrounds the

instrument. The instrument sits on the base of the shipping container, as shown in Figure 3.

C

M

Y

CM

MY

CY

CMY

K

AriaMx_Poster_18x12_2.6.pdf   1   1/15/15   8:56 AM

Figure 5 

Optical module housing 

carrier,  centered  in  the  instru

-

ment door opening