Performance Test Record
Table 2-2, "Performance Test Record," is a form you may copy and use to
record performance verification test results for the multiplexers. Table 2-2
shows multiplexer test limits, DMM measurement uncertainty, and test
accuracy ratio (TAR) values.
Test Limits
Test limits are defined for relay closed channel resistance and DC isolation
using the specifications in the individual module’s User’s Manual. The relay
contact resistance and DC isolation specifications are single-ended, meaning
that there is an upper limit OR a lower limit, but not both. In Table 2-2, the
Minimum or Maximum column is blank for a single-sided test.
Measurement
Uncertainty
For the performance verification tests in this manual, measurement
uncertainties are calculated based on the 3458A Digital Multimeter.
The measurement uncertainty shown in Table 2-2 is the accuracy of the
3458A using 90-day specifications. The calculations follow.
Closed Channel
Resistance Test
Conditions:
•
4-wire ohms function, 10
Ω
range
•
90-day specifications
•
Worst-case reading = 2.0
Ω
M.U. = (15
ppm of Reading
+ 5
ppm of Range
)
=
(
15
×
10
−
6
⋅
2.0
+
5
×
10
−
6
⋅
10
)
(Ω)
=
8.0
×
10
−
5
Ω
__________
DC Isolation Test
Conditions:
•
2-wire ohms function, 1 G
Ω
range
•
90-day specifications
•
Worst-case reading = 1.2 G
Ω
(highest resistance that can
be measured with the 3458A)
M.U. = (0.5%
of Reading
+ 10
ppm of Range
)
=
(
0.005
⋅
1.2
×
10
9
+
10
×
10
−
6
⋅
1
×
10
9
)
(Ω)
=
6.0
×
10
6
Ω
_________
Test Accuracy Ratio
(TAR)
Test Accuracy Ratios (TAR) are not defined for single-sided measurements,
so all closed-channel resistance and DC isolation measurements show NA
(Not Applicable) in the TAR column.
Chapter 2
Verification Tests 37
Summary of Contents for Agilent E1345A
Page 2: ......
Page 8: ...Notes 6 E1343A 44A 45A 47A 55A 56A Service Manual ...
Page 9: ......
Page 10: ......
Page 12: ...Notes 10 What s In This Manual E1343A 44A 45A 47A 55A 56A Service Manual ...
Page 20: ...Figure 1 2 Initial Incoming Inspection Guidelines 18 General Information Chapter 1 ...
Page 22: ...Notes 20 General Information Chapter 1 ...
Page 50: ...Figure 3 3 E1355A E1356A Terminal Module Mechanical Parts 48 Replaceable Parts Chapter 3 ...
Page 54: ...Figure 4 1 E1343A E1344A E1345A E1347A Multiplexer Block Diagram 52 Service Chapter 4 ...
Page 56: ...Figure 4 2 E1355A 56A Multiplexers Block Diagram 54 Service Chapter 4 ...
Page 64: ...Notes 62 Service Chapter 4 ...
Page 72: ...Notes 70 Example C Programs Appendix A ...
Page 82: ...Figure B 3 E1355A 56A Terminal Module Mechanical Parts 80 Backdating Information Appendix B ...
Page 83: ......
Page 84: ...Copyright Agilent Technologies Inc 2005 E1345 90013 E1345 90013 ...