Read
Error-Corrected
Data
230
!
Read
Error-Corrected
Data
240
DIM
Dat(1:801,1:2)
!
When
iBASIC
is
used,
change
"801"
to
"802"
250
OUTPUT
@Hp4396;"POIN?"
!
\
260
ENTER
@Hp4396;Nop
!
|
When
iBASIC
is
used,
delete
these
lines
270
REDIM
Dat(1:Nop,1:2)
!
/
280
ASSIGN
@Dt
TO
717;FORMAT
OFF
!
When
iBASIC
is
used,
290
OUTPUT
@Hp4396;"FORM3"
!
change
"717"
to
"800"
300
OUTPUT
@Hp4396;"OUTPDATA?"
310
ENTER
@Dt
USING
"%,8A";Head$
320
ENTER
@Dt;Dat(*)
330
ENTER
@Dt
USING
"%,1A";Dum$
!
When
iBASIC
is
used,
delete
this
line
OUTPDATA?
command
(line
300)
retrieves
D
A
T
A
ARRA
YS
in
the
analyzer
.
F
or
details
on
data
transfer
,
see
Chapter
4.
Restore
Modied
Error-Corrected
Data
370
!
Restore
Error-Corrected
Data
380
OUTPUT
@Hp4396;"INPUDATA
";
390
OUTPUT
@Dt
USING
"#,8A";Head$
400
OUTPUT
@Dt;Dat(*),END
Line
380
opens
the
D
A
T
A
ARRA
YS
in
the
analyzer
to
restore
the
data.
Lines
390
to
400
transfer
data
in
FORM3
(a
similar
procedure
is
used
in
the
\T
o
Modify
Calibration
Data
"
example).
Writing
Data
Arrays
to
the
Analyzer
5-7