Configure a Ramp Wave
Configure a Ramp Wave
Description
A ramp wave has amplitude, offset, and phase relative to sync pulse. It also has symmetry for creating triangular and
other similar waveforms. Its amplitude and offset can also be set using high and low voltage values.
Example
The following waveform can be set up with the series of SCPI commands, where high and low can be used in place of
and
The following commands produce the ramp wave shown above.
FUNCtion RAMP
FUNCtion:RAMP:SY50.0
FREQ +1.0E+03
V2.0
VOLTage:HIGH +2.0
VOLTage:LOW +0.0
VOLTage:1.0
OUTPut1 1
Remarks
l
Ramp frequency is limited to 200 kHz.
l
Although period can be adjusted from the instrument's front panel, there is no SOUR:FUNC:RAMP:PER or
SOUR:PER command that can be used in addition to
.
330