Controller to Pump Connection
A five-meter long cable is to connect the controller to the pump. The
following figures show the controller output connector configuration
where pins:
A-F = upper bearing temperature sensor
A-G = lower bearing temperature sensor
B-C-D = 120 Vac 3-phase output to pump motor stator
E = ground
Figure 11
Controller-to-pump connector (applicable to model 969-9462 and
969-9562)
Turbo-V 2000HT Controller User Manual / 87-900-913-01 (O)
249/306