Technical Information
Procedure to Connect the Serial and I/O ports
to an External Cable
288/306
Turbo-V 903 Pumping System User Manual
/ 87-901-016-01
(A)
Window-Meanings
Table 6
Window-Meanings
N. Read/Write
Datatype
Description
Admitted
Values
000
R/W
L
Start/Stop (in remote mode
the window is a read only)
Start = 1
Stop = 0
008 R/W
L
Remote
(default)
or Serial configuration
Remote = 1
Serial = 0
(default = 1)
100
R/W
L
Soft Start (write only in Stop
condition)
YES = 1
NO = 0
101
R/W
N
Set Point type
0 = Frequency
1 = Current
2 = Time
(default = 0)
102
R/W
N
Set Point threshold
(expressed in Hz, mA or s)
(default = 867)
103
R/W
N
Set Point delay: time
between the pump start and
the set point check (seconds)
0 to 99999
(default = 0)
104
R/W
L
Set Point signal activation
type: the signal can be "high
level active" or "low level
active"
0 = high level
active
1 = low level
active
(default = 0)
105
R/W
N
Set point hysteresis
(in % of threshold)
0 to 100
(default = 2)
106
R/W
L
Water cooling
0 = NO
1 = YES
107 R/W
L
Active
Stop
(write only in stop)
0 = NO
1 = YES
108
R/W
N
Baud rate
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
Summary of Contents for 969-6500
Page 4: ...Turbo V 903 Pumping System 4 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Page 14: ...Contents 14 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Page 58: ...3 Mode d emploi Mise au rebut 58 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Page 118: ...7 Istruktionsbog Bortskaffelse 118 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Page 176: ...11 Oδηγιεσ Χρησεωσ Χώνευση 176 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Page 218: ...14 Návod k Použití Likvidace 218 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Page 232: ...15 Návod na Obsluhu Likvidácia 232 306 Turbo V 903 Pumping System User Manual 87 901 016 01 A ...
Page 301: ......
Page 302: ......
Page 303: ......
Page 304: ......
Page 305: ......