To Clean the Septum Seat in the Insert Assembly of the MMI
132
Maintaining Your GC
To Clean the Septum Seat in the Insert Assembly of
the MMI
1
Gather the following:
•
Replacement septum, see
“Consumables and Parts for the Multimode Inlet”
•
Wrench, hex for changing septum
•
0- or 00-grade steel wool (optional)
•
Tweezers
•
Compressed, filtered, dry air or nitrogen
2
Manually set the inlet and oven temperature to < 40
°
C, and wait for the inlet, oven, and
other parts you might come into contact with inside the oven, to cool before continuing.
Alternately, place the GC in maintenance mode:
Maintenance > Instrument > Perform
Maintenance > Maintenance Mode > Start Maintenance.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands
.
3
Slide the locking tab forward (counterclockwise). Lift the insert assembly straight up and
away from the inlet to avoid chipping or breaking the liner.
4
Remove the septum retainer nut or Merlin cap.
5
Use tweezers to remove the septum or Merlin Microseal from the retainer nut. See
6
Scrub the residue from the retainer nut and septum holder with a small piece of rolled-up
steel wool and tweezers. Do not do this over the inlet.
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...