Maintaining the COC Inlet
To Check the Needle-to-Column Size on the COC Inlet
Maintaining Your GC
109
To Check the Needle-to-Column Size on the COC
Inlet
1
Gather the following:
•
Insert, see
“Consumables and Parts for the Cool On-Column Inlet”
•
Syringe needle
2
Place GC in maintenance mode:
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance.
Wait for the GC to become ready.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands
.
WARNING
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
3
Check that the needle is the correct size for the desired column id. If needed, replaced the
needle with one of the correct size. See
“Consumables and Parts for the Cool On-Column
“To Replace a Needle in a Syringe”
the Fused Silica Needle in a Syringe for the COC Inlet”
4
Identify the correct insert for the column size. See
“Consumables and Parts for the Cool
on page 102. Use the insert that is the same size as the syringe needle to
verify that the column you plan to use is the correct size.
5
Make a clean cut on the end of the column. See
“To Install a Capillary Column with the
6
Insert the column into one end of the insert.
7
Insert the syringe needle through the other end of the insert and into the column. The
needle should visibly enter the column without any obstruction. If the needle cannot pass
easily into the column, reverse the insert to try the needle and column in the other end.
8
Exit maintenance mode. Select
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
Syringe
Insert
Column
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...