To Replace a Needle in a Syringe
Maintaining Your GC
117
To Replace a Needle in a Syringe
1
Gather the following, see
, “Recommended parts for injections onto 0.25-mm
fused silica columns,” on page 103:
•
Syringe barrel
•
Needle, 250-µm or 320-µm
2
Unscrew the syringe barrel cap and remove the spring.
3
Make sure that the needle has the PTFE disk as shown below. If the syringe barrel does not
have the PTFE disk, use the instructions in the syringe box to wrap the needle.
4
Check the new needle for a small wire inserted for shipment. Remove the wire if present.
5
Slide the spring and the cap over the needle.
6
Insert the needle into the syringe barrel.
7
Screw the cap back on the syringe barrel.
Plunger
Barrel
Needle
Spring
Cap
PTFE disk
Needle
Stop
Summary of Contents for 8890
Page 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Page 8: ...8 Maintaining Your GC ...
Page 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Page 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Page 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Page 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Page 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Page 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Page 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...