5-6
Automated Verification – Option 011
Step 2. Measurement Calibration
Step 2. Measurement Calibration
1
Press
CAL
,
CAL KIT & STDS
,
SELECT CAL KIT
,
CAL KIT:7mm
,
RETURN
,
RETURN
,
CALIBRATE
MENU
,
FULL 2-PORT
.
2
Press
ISOLATION
,
OMIT ISOLATION
,
ISOLATION DONE
.
3
Press
REFLECTION
.
4
Connect the “open” end of the open/short combination (supplied in the
calibration kit) to reference test port 1, as shown in
.
Figure 5-2. Connections for Measurement Calibration Standards
5
Press
FORWARD:OPEN
. The analyzer displays the message
WAIT--MEASURING
CAL STD
.
6
When the analyzer finishes measuring the standard, connect the “short” end of
the open/short combination to reference test port 1, as shown in
Figure 5-2
.