![Agilent Technologies 8702D Installation Manual Download Page 113](http://html.mh-extra.com/html/agilent-technologies/8702d/8702d_installation-manual_2867953113.webp)
4-57
Performance Tests
15. Test Port Receiver Magnitude Compression
7
For every test port input power level, calculate the receiver phase dynamic
accuracy using the following formula:
Phase Dynamic Accuracy = SIN
–1
(B)
8
Write the value in the “Results Value” column of the “Performance Test
Record.”
Test Port 1 Phase Dynamic Accuracy
9
through
.
15. Test Port Receiver Magnitude Compression
Perform this test to verify the compression/expansion magnitude levels of the
analyzer’s test port receiver samplers.
Required
Equipment
For 50
Ω
Analyzers
Cable, APC-7, 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4779
For 75
Ω
Analyzers
Cable, 75
Ω
, Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-2408
Procedure
Test Port 2 Magnitude Compression
1
Connect the equipment as shown in
.
Figure 4-34. Test Port Magnitude Compression Test Setup