Chapter 7
333
General Troubleshooting
Error Messages
Sampling Oscillator (355)
This error indicates an unlocked sampling oscillator
during the local oscillator (LO) alignment routine. This
error only applies to the hardware in an 8560 E-series
or EC-series spectrum analyzer.
355 SMP CAL
Sampler unlock condition during calibration routine.
This error remains until a successful recalibration is
performed.
Span Accuracy Calibration Errors (356 to 361)
These errors are generated when the span accuracy
calibration fails. The span accuracy calibration is done
during "power up", IF calibration (every 5 minutes),
and LO IF realignment routines. Span accuracy
calibration sweeps occur during the retrace (dead time)
of the main sweep ramp. The firmware then detects any
span accuracy calibration errors. These errors only
apply to firmware revisions 931216 and later.
356 SPAC CAL
Sweep data problem finding "bucket 1" of the span
accuracy calibration sweep. This error indicates a
possible failure of the sweep generator, span
attenuator, or main/FM coil sweep switches on the A14
frequency control assembly. Refer to
357 SPAC CAL
Cannot find the "x" intersection for "bucket 1" of the
span accuracy calibration sweep. This error indicates a
possible failure of the sweep generator, span
attenuator, or main/FM coil sweep switches on the A14
frequency control assembly. Refer to
358 SPAC CAL
Sweep data problem finding "bucket 2" of the span
accuracy calibration sweep. This error indicates a
possible failure of the sweep generator, span
attenuator, or main/FM coil sweep switches on the A14
frequency control assembly. Refer to
359 SPAC CAL
Cannot find "x" intersection for "bucket 2" of the span
accuracy calibration sweep. This error indicates a
possible failure of the sweep generator, span
attenuator, or main/FM coil sweep switches on the A14
frequency control assembly. Refer to
Summary of Contents for 8564EC
Page 17: ...25 1 General Information ...
Page 37: ...47 2 Adjustment Diagnostic Software ...
Page 77: ...89 3 Manual Adjustment Procedures ...
Page 129: ...161 3a Manual Adjustment Procedures 3335A Source not Available ...
Page 142: ...175 4 Assembly Replacement ...
Page 194: ...Chapter 4 257 Assembly Replacement Procedure 13 A21 OCXO Figure 4 34 A21 OCXO Mounting Screws ...
Page 196: ...259 5 Replaceable Parts ...
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ...303 6 Major Assembly and Cable Locations ...
Page 234: ...315 7 General Troubleshooting ...
Page 238: ...Chapter 7 319 General Troubleshooting Introduction Figure 7 2 Ribbon Cable Connections 1 of 3 ...
Page 239: ...320 Chapter7 General Troubleshooting Introduction Figure 7 3 Ribbon Cable Connections 2 of 3 ...
Page 242: ...Chapter 7 323 General Troubleshooting Introduction Figure 7 5 Service Cal Data Menu ...
Page 271: ...352 Chapter7 General Troubleshooting Block Diagram Description Figure 7 6 Functional Sections ...
Page 283: ...364 Chapter7 General Troubleshooting Block Diagram Description ...
Page 284: ......
Page 285: ......
Page 286: ......
Page 287: ...377 8 ADC Interface Section ...
Page 291: ...Chapter 8 381 ADC Interface Section Introduction Figure 8 2 A3 Test Connector Pin Locations ...
Page 321: ......
Page 322: ...427 9 IF Section ...
Page 356: ...Chapter 9 461 IF Section A5 IF Assembly Figure 9 13 Detailed IF Adjust Signature 5 ...
Page 376: ......
Page 377: ......
Page 378: ......
Page 379: ...487 10 Controller Section ...
Page 394: ...521 11 Synthesizer Section ...
Page 451: ......
Page 452: ......
Page 453: ......
Page 454: ...593 12 RF Section ...
Page 489: ...628 Chapter12 RF Section A15 RF Assembly Figure 12 10 10 MHz TTL Reference at U304 Pin 13 ...
Page 491: ......
Page 492: ...633 13 Display Power Supply Section ...
Page 504: ......
Page 505: ...671 14 Component Level Information Packets ...
Page 507: ......