Agilent Technologies 85130C Operating And Service Manual Download Page 3

85130C Operating and Service Manual

 1

85130C Adapter Kit

General Information

To obtain optimum performance from this adapter kit, observe these simple precautions:

Make connections carefully to avoid misalignment and connector damage, which will result in 
inaccurate measurements.

Keep the connectors free of dirt and any particles.

When you clean the connectors, try using compressed air first. Do not use abrasives. With a clean foam 
swab, apply only isopropyl alcohol.

For more information, refer to the 

Connector Care for RF and Microwave Coaxial Connectors 

document. It 

can be viewed online by searching for part number 08510-90064 at 

www.agilent.com

.

Description

The 85130C adapters are used when type-N devices must be attached to a 3.5 mm test set in a network 
analyzer system. The test set end of the adapters has a NMD-3.5 mm (f) connector while the Device Under 
Test (DUT) end has a precision slotless type-N connector (PSC-N).

Contents

The 85130C kit contains the following:

Test port adapter, NMD-3.5 mm (f) to PSC-N (m) (Agilent part number 85054-60030)

Test port adapter, NMD-3.5 mm (f) to PSC-N (f) (Agilent part number 85054-60029)

Storage box, foam-lined

Operating  and  Service  Manual

Spanner wrench

Summary of Contents for 85130C

Page 1: ...Agilent Technologies Agilent 85130C NMD 3 5 mm to Type N Adapter Kit Operating and Service Manual...

Page 2: ...ument are furnished under a license and may be used or copied only in accordance with the terms of such license Restricted Rights Legend If software is for use in the performance of a U S Government p...

Page 3: ...o the Connector Care for RF and Microwave Coaxial Connectors document It can be viewed online by searching for part number 08510 90064 at www agilent com Description The 85130C adapters are used when...

Page 4: ...e male connector See Figure 1 on page 3 Zero the gages before each use by attaching the appropriate gage master to the end of the gage torquing the connection to 12 in lb with a 3 4 inch torque wrench...

Page 5: ...dically repeat the test to determine if the performance meets the electrical specifications on the previous page or if they need to be replaced An initial period of one year between performance tests...

Page 6: ...functions mentioned in the return loss test below 1 Preset the analyzer 2 Set a stimulus start frequency of the analyzer s lowest frequency 3 Set a stimulus stop frequency of 18 GHz 4 Set an IF bandw...

Page 7: ...laceable Parts There are no replaceable components for the adapters A worn or damaged adapter must be replaced in whole ELECTRICAL SPECIFICATIONS Tested by ______________ Date ______________ Adapter F...

Page 8: ...espondence or telephone conversation refer to the Agilent product by its model number and full serial number With this information the Agilent representative can determine whether your product is stil...