General Information
Calibration Kit Overview
Calibration Kit Overview
The Agilent 85056D 2.4 mm calibration kit is used to calibrate Agilent network analyzers
up to 50 GHz for measurements of components with 2.4 mm connectors.
Kit Contents
The 85056D calibration kit includes the following items:
• offset opens and shorts
• broadband terminations
• 2.4 mm adapters
• 5/16 in, 90 N-cm (8 in-lb) torque wrench
• 7 mm open-end wrench
Refer to Chapter 6 , “Replaceable Parts,” for a complete list of kit contents and their
associated part numbers.
NOTE
85056D kits with serial prefix 3012A did not have a calibration definitions
disk to support the 8510C network analyzer. The part numbers provided in
this manual are the recommended replacement parts for these kits. The
devices in these kits should meet the specifications published in this manual.
Broadband Loads
The broadband loads are metrology-grade, 50
Ω
terminations that have been optimized for
performance up to 50 GHz. The rugged internal structure provides for highly repeatable
connections. A distributed resistive element on sapphire provides excellent stability and
return loss.
Offset Opens and Shorts
The offset opens and shorts are built from parts that are machined to the current
state-of-the-art in precision machining.
The offset short’s inner conductors have a one-piece construction, common with the
shorting plane. The construction provides for extremely repeatable connections.
The offset opens have inner conductors that are supported by a strong,
low-dielectric-constant plastic to minimize compensation values.
Both the opens and shorts are constructed so that the pin depth can be controlled very
tightly, thereby minimizing phase errors. The lengths of the offsets in the opens and shorts
are designed so that the difference in phase of their reflection coefficients is approximately
180 degrees at all frequencies.
Adapters
Like the other devices in the kit, the adapters are built to very tight tolerances to provide
Summary of Contents for 85056D
Page 4: ...iv 85056D...
Page 6: ...vi 85056D Contents...
Page 7: ...1 1 1 General Information...
Page 15: ...2 1 2 Specifications...
Page 20: ...2 6 85056D Specifications Electrical Specifications...
Page 21: ...3 1 3 Use Maintenance and Care of the Devices...
Page 34: ...3 14 85056D Use Maintenance and Care of the Devices Handling and Storage...
Page 38: ...4 4 85056D Performance Verification Recertification...
Page 39: ...5 1 5 Troubleshooting...
Page 44: ...5 6 85056D Troubleshooting Contacting Agilent...
Page 45: ...6 1 6 Replaceable Parts...
Page 48: ...6 4 85056D Replaceable Parts Figure 6 1 Replaceable Parts for the 85056D Calibration Kit...
Page 49: ...A 1 A Standard Definitions...