A-2
85052D
Standard Definitions
Standard Class Assignments
Standard Class Assignments
Class assignment organizes calibration standards into a format compatible with the error
models used in the measurement calibration. A class or group of classes corresponds to the
systematic errors to be removed from the measured network analyzer response.
through
list the classes of the devices in the kit for various network analyzers. This
information resides on the calibration constants disk included in the kit.
Table A-1
Standard Class Assignments for the 8510 Network Analyzer
Disk File Name: CK_35mmD3
Calibration Kit Label:
3.5 mm D.3
Class
A
B
C
D
E
F
G
Standard Class Label
S
11
A
2
Open
S
11
B
1
Short
S
11
C
9
Loads
S
22
A
2
Open
S
22
B
1
Short
S
22
C
9
Loads
Forward transmission
11
Thru
Reverse transmission
11
Thru
Forward match
11
Thru
Reverse match
11
Thru
Forward isolation
a
a. The forward isolation standard is also used for the isolation part of the response and isolation calibration.
9
Isol’n Std
Reverse isolation
9
Isol’n Std
Frequency response
1
2
11
Response
TRL thru
Undefined
TRL reflect
Undefined
TRL line
Undefined
Adapter
13
5
6
7
8
Adapter
TRL Option
Cal Z
0
: _____ System Z
0
__X__ Line Z
0
Set ref: __X__ Thru
_____ Reflect
Lowband frequency: ________
Summary of Contents for 85052D
Page 4: ...iv...
Page 7: ...85052D 1 1 1 General Information...
Page 13: ...85052D 2 1 2 Specifications...
Page 19: ...85052D 3 1 3 Use Maintenance and Care of the Devices...
Page 31: ...85052D 4 1 4 Performance Verification...
Page 34: ...4 4 85052D Performance Verification Recertification...
Page 35: ...85052D 5 1 5 Troubleshooting...
Page 39: ...85052D 6 1 6 Replaceable Parts...
Page 44: ...6 6 85052D Replaceable Parts Introduction...
Page 45: ...85052D A 1 A Standard Definitions...
Page 62: ...Index iv 85052D Index...