![Agilent Technologies 85024A User'S And Service Manual Download Page 33](http://html.mh-extra.com/html/agilent-technologies/85024a/85024a_users-and-service-manual_2867925033.webp)
Chapter 5
5-3
Performance Tests
Recommended Test Equipment
Recommended Test Equipment
lists the equipment that is recommended for use in performance testing of the probe. Other
equipment may be substituted if its specifications meet or exceed the specifications listed in the “Critical
Specifications” column.
Table 5-1 Recommended Test Equipment
Type
Critical Specifications
Recommended Model or Part
Number
Vector Network Analyzer
300 kHz to 3 GHz
8714ET/ES, 8753ET/ES, 8753E
a
a. Many models of network analyzers can be used for these performance tests. The network analyzer
must cover the frequency range 300 kHz to 3 GHz, and be capable of power output in the range of
3
to +5 dBm.
Power Meter
437B/438A, or E4418A/E4419A
Power Sensor
300 kHz to 3 GHz
30 dBm to +5 dBm
8482A
Type-N Cable, male-to-male
50
8120-8862 or equivalent
Probe Adapter
Supplied with probe
11880-60001
Attenuator, type-N, 50
any value 10 to 20 dB
nominal
8491A/B/C Option 20, Option 10, or
equivalent
Type-N Adapter, female-to-female
1250-0777 or equivalent
Summary of Contents for 85024A
Page 4: ...iv ...
Page 7: ...1 1 1 General Information ...
Page 13: ...2 1 2 Accessories ...
Page 19: ...3 1 3 Installation ...
Page 24: ...3 6 Chapter3 Installation Returning the Product for Service ...
Page 25: ...4 1 4 Operation ...
Page 30: ...4 6 Chapter4 Operation Operator s Check ...
Page 31: ...5 1 5 Performance Tests ...
Page 40: ...5 10 Chapter5 Performance Tests Average Noise Level ...
Page 41: ...6 1 6 Replaceable Parts ...
Page 46: ...6 6 Chapter6 Replaceable Parts Parts Lists ...
Page 47: ...7 1 7 Service ...
Page 60: ...7 14 Chapter7 Service Replacement Procedure Figure 7 11 Regulator Parts and Wiring ...
Page 64: ...7 18 Chapter7 Service Connector Inspection and Cleaning ...