3-10
Calibration Overview
User Calibrations—Optical and Electrical
To perform an O/E user-wavelength calibration
1
Press the plug-in module’s front-panel optical channel
SETUP
key.
2
Press
Calibrate
, and then
O/E cal
.
3
Input the correct wavelength, and follow the instructions on the screen.
Figure 3-2. Plug-in calibration menu
To use an O/E user-wavelength calibration
1
Press the plug-in module’s front-panel optical channel
SETUP
key.
2
Press
Bandwidth/wavelength
and then
wavelength
.
3
Press
Usr
wavelength
and then
Enter
.
NOTE
The optical channel calibration accuracy is heavily dependent on the accuracy to which
you know the optical source power. For best results, measure the optical source power
with an optical power meter such as the Agilent 8153A and use precision optical con-
nectors. In addition, proper connector cleaning procedures are essential to obtaining an
accurate calibration.
Summary of Contents for 83487A
Page 1: ...Agilent 83487A Optical Electrical Plug In Module User s Guide ...
Page 8: ...viii General Safety Considerations ...
Page 10: ......
Page 25: ...2 Channel Setup Menu 2 2 Displaying the Channel Setup Menus 2 5 Channel Setup Menu ...
Page 27: ...2 3 Channel Setup Menu Channel Setup Menu Figure 2 1 Optical Channel Setup menu ...
Page 28: ...2 4 Channel Setup Menu Channel Setup Menu Figure 2 2 Electrical Channel Setup menu ...
Page 38: ......
Page 68: ......
Page 96: ......