4-6
Performance Tests
Tests for Sources
3. Electrical Input Port Match
1
Connect the equipment as shown in
.
Figure 4-3. Electrical input port match test setup
2
Turn on the equipment, and allow them to warm up for 1 hour.
3
Use the Agilent 8702D’s “Guided Setup” feature to perform an electrical-to-
electrical response test. Use the following settings:
Start frequency: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300 kHz
Stop frequency (Agilent 83403C): . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 GHz
Stop frequency (Agilent 83402C): . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 GHz
Stimulus power: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 dBm
IF bandwidth: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 Hz
4
The guided setup routine will step you through an S11, 1-port, electrical
calibration.
5
Place the Agilent 8702D in single-sweep mode, and use the marker functions
to locate the peak of the response.
6
Compare the measurement with the specification listed in
“Specifications and Regulatory Information”
.