Performance Tests
Performance Test Instructions
74
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
6
At the Return Loss module,
a
Make sure [
FP Delta
] is set to 0.00 dB
b
Make sure [
RL ref
] is set to 0.00 dB
c
Set the averaging time [
Av Time
] to 10 s
d
Set <
Source wavelenth
> to External.
7
Turn on the 11896A Polarization Controller.
a
Set the ‘Scan Rate’ to 8
b
Press the ‘Autoscan’ button to begin scrambling the polarization
states.
8
Enable the Attenuator, and set the ‘Attenuation’ such that the
Power Sensor shows a power [
P
] reading of exactly -20.0000 dB
9
Moving the cable as little as possible, disconnect the 81610CC
Reference Cable from the Power Sensor module, and connect it to
the output of the Return Loss module as shown in Figure 43
Figure 43
81610CC Reference Cable - Setting the Calibration Reference
10
At the Return Loss module, press [RefCal].
11
Disconnect the 81610CC Reference Cable from the Attenuator, and
connect the straight end to the Power Sensor module, as shown in
Use an additional 81113PC patchcord to connect the the straight
Attenuator output to the angled Return Loss module input.
81610CC
Optical Attenuator 8156A
8156A #101
Mainframe 8163A w/ 8161xA RTL
and 81634A Power Sensor
8163A
8161xA
in
out
81634A
Mainframe 8163A w/ 81654A FP
Laser Source
8163A
81654A
Polarization Ctrl 11896A
11896A
Summary of Contents for 81610A
Page 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Page 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Page 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Page 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Page 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Page 116: ......