Performance Test Instructions
Performance Tests
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
73
Calibrating the 81610CC Reference Cable
The 81610CC Reference Cable may be recalibrated using this
procedure. Repair is not possible.
1
Make sure that all cable connectors are clean.
2
Setup the equipment as shown in Figure 42.
Make sure that the ouptut port of the Attenuator has a straight
connector. Clean this connector very carefully, and connect it to the
straight end of the 81610CC Reference Cable.
Figure 42
81610CC Reference Cable - Initial Setup
3
At the Laser Source module,
a
Enable the Laser Source
b
Set the wavelength [
λ]
to 1310 nm
c
Press [
Mod Src
] and select <
Coherence Ctrl
> to modulate the
source using coherence control.
N O T E
Allow the Laser Source to stabilize for at least 20 minutes.
4
At the Attenuator, set the ‘Wavelength’ to the wavelength of the
Laser source.
5
At the Power Sensor module,
a
Set the wavelength [
λ]
to the wavelength of the Laser Source.
b
Set [Pwr unit] to <
dB
>.
81610CC
Optical Attenuator 8156A
8156A #101
Mainframe 8163A w/ 8161xA RTL
and 81634A Power Sensor
8163A
8161xA
in
out
81634A
Mainframe 8163A w/ 81654A FP
Laser Source
8163A
81654A
Polarization Ctrl 11896A
11896A
Summary of Contents for 81610A
Page 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Page 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Page 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Page 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Page 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Page 116: ......