Making a Return Loss Measurement
Getting Started with Return Loss
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
31
How to Measure Return Loss
It is not necessary to make new calibration measurements for each
DUT. You can make the calibration measurements for your system,
and then measure the return loss of many devices.
The value shown in the result field for the Return Loss channel is the
measured return loss.
1
Attach the DUT to the measurement patchcord. In the example
shown in Figure 24, the DUT is a connector pair.
Figure 24
Measuring the Return Loss of the DUT (in this example: a Connector Pair)
T I P
Terminate your system close to the DUT to make sure that you are
only measuring reflections from the DUT.
How to Measure Return Loss and Insertion Loss
The return loss module together with an additional power head allow
you to perform combined return loss and insertion loss
measurements. The figure below shows the standard measurement
setup for the reference measurement and the device measurement.
Figure 25
Measuring the Power Transmitted through the DUT (a Connector Pair)
Termination
measurement patchcord
8161x
8163A Lightwave Multimeter
Laser
Source
Agilent 81113PC
Power
Sensor
8163A Lightwave Multimeter
measurement patchcord
8161x
8163A Lightwave Multimeter
Laser
Source
Agilent 81113PC
Summary of Contents for 81610A
Page 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Page 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Page 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Page 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Page 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Page 116: ......