Making a Return Loss Measurement
Getting Started with Return Loss
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
23
Calibration using the Agilent 81000BR
Reference Reflector
You can use the Agilent 81000BR Reference Reflector to make a
Reflectance Calibration, and a Termination Calibration.
Reflectance Calibration
Use this procedure to calibrate the Return Loss module against a
component of known return loss.
1
Making sure all the connectors are clean, set up the instrument as
shown in Figure 11 if you are using an External Source,
Figure 11
Reflectance Calibration - External Source
or Figure 12 if you are using an internal source
(Agilent 81611A/2A/3A/4A Return Loss modules only).
Figure 12
Reflection Calibration - Internal Source
T I P
For best results and higher repeatability, fix the cable.
2
Attach a component with a known return loss to the end of the
patchcord.
T I P
The Agilent 81000BR Back Reflector is such a component, offering a
return loss of 0.18
±
0.1 dB.
3
Move to the Return Loss channel, then press [Details]
4
Move to [
FPDelta
], press [Edit]
,
set the value to 0.000, then press
[OK].
Laser
Agilent 81113PC
Agilent 81000BR Agilent 81000FI
Agilent 81000UM
8161x
8163A Lightwave Multimeter
Laser
Source
Agilent 81113PC
8161x
8163A Lightwave Multimeter
Agilent 81113PC
Agilent 81000BR
Agilent 81000FI
Agilent 81000UM
8161x
8163A Lightwave Multimeter
Summary of Contents for 81610A
Page 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Page 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Page 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Page 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Page 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Page 116: ......