Calibration Measurements
Getting Started with Return Loss
Agilent 81610A/11A/12A/13A/14A Return Loss Module, Fifth Edition
19
Calibration Measurements
Before measuring the reflection factor of a device under test (DUT),
take calibration measurements as described in “Calibrating the Return
Loss Module” on page 21. These calibrations eliminate wavelength
dependencies, coupler directivity, insertion losses, backscattering and
other non-ideal characteristics of the system.
Making a Return Loss
Measurement
Connectors
The Return Loss measurement setup described uses Diamond HMS-
10/Agilent/HRL and Diamond HMS-10/Agilent connectors throughout.
Patchcord on RL module Output
It is recommended that you attach a patchcord with a high return loss
connector to the Return Loss module output.
C A U T I O N
Agilent Technologies supplies patchcords with a Diamond HMS-
10/Agilent/HRL high return loss connector on one end. These
patchcords are necessary so that the connector at the output is not
damaged. The full range of patchcords available are described in Table
Table 2
High Return-Loss Patchcords
Model No.
High Return-Loss Patchcords
Description
Agilent 81113EC
DIN47256/4108 (angled) - Radiall EC
Agilent 81113PC
DIN47256/4108 (angled) - FC/PC
Agilent 81113SC
DIN47256/4108 (angled) - DIN47256/4108 (angled)
Agilent 81113BC
DIN47256/4108 (angled) - Bare Fiber
Summary of Contents for 81610A
Page 2: ...S S1 Agilent 81610A 11A 12A 13A 14A Return Loss Module User s Guide ...
Page 48: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 45 Accessories ...
Page 52: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 49 Specifications ...
Page 62: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 59 Performance Tests ...
Page 92: ...Agilent 81610A 11A 12A 13A 14A Return Loss Module Fifth Edition 89 Cleaning Information ...
Page 116: ......