6 Changing the ParBERT Hardware
Installing a Module
6-4
Agilent 81250 ParBERT Installation Guide, May 2004
If you want to replace the module, insert the new module now, as
described under
“Installing a Module” on page 6-4
. If not, remember that
a clockgroup does not tolerate empty slots. Data modules following an
empty slot will be ignored by the system.
Installing a Module
To install a module:
1 Switch off the ParBERT mainframe.
2 Choose an empty mainframe slot.
A single ParBERT system does not tolerate empty slots.
All data modules of a system have to be in adjacent slots at the right-hand
side of the ckock module. Data modules following an empty slot are
ignored.
If you wish to use multiple independent ParBERT systems, you can install
more than one clock module in one mainframe, each followed by its data
modules. In this case, you may leave empty slots before the additional
clock modules.
3 Set the module’s logical address.
4 Plug the module into the mainframe.
5 Connect the module.
Refer to the following sections.
Overview
NOTE
Details
Summary of Contents for 81250
Page 1: ...Agilent Parallel Bit Error Ratio Tester 81250 ParBERT Installation Guide Agilent Technologies ...
Page 12: ...1 About this Manual Document History 1 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 30: ...2 Introduction Unpacking ParBERT 2 18 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 139: ...Declaration of Conformity Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 3 ...
Page 141: ...Site Attenuation Requirements Appendix Agilent 81250 ParBERT Installation Guide May 2004 10 5 ...
Page 142: ...Appendix Site Attenuation Requirements 10 6 Agilent 81250 ParBERT Installation Guide May 2004 ...
Page 145: ......