Introduction
Oven thermal leaks
Do not allow wiring or temperature probes to pass through the oven door jam.
Agilent recommends using one of the access holes.
WA R N I N G
The insulation around the inlets, detectors, valve box, and the insulation cups is
made of refractory ceramic fibers. To avoid inhaling fiber particles, we recommend
the following safety procedures: ventilate your work area; wear long sleeves,
gloves, safety glasses, and a disposable dust/mist respirator; dispose of insulation
in a sealed plastic bag; wash your hands with mild soap and cold water after
handling the insulation.
WA R N I N G
Objects passing through the oven door seal can cause thermal leaks which create
hazardous hot spots which cause burns and melt equipment.