3-69
Eye-Diagram Analyzer Reference
Agilent 70820A Calibration
5
To view the data for channel 1, press:
CH1 CAL INFO
6
To view the data for channel 2, press:
CH2 CAL INFO
7
To view additional pages of information, continue pressing:
CH1 CAL INFO
or
CH2 CAL INFO
Performing a Self-Test
You can initiate an automatic self-test routine that checks the internal condi-
tion of all major circuits. The self-test takes approximately 30 seconds to run.
(This routine automatically runs whenever the microwave transition analyzer
is turned on.)
Perform the self-test by pressing:
MENU
,
page 1 of 2
,
Calib
,
SELF-TEST
Controlling the Calibrator Signal
For verification purposes, the Calibrator Output signal can be manually con-
trolled. This includes changing its amplitude and frequency. The signal has the
following characteristics:
The signal’s amplitude can be set to one of five positions: 0 through 4. Zero
represents the smallest available amplitude. Four represents the largest
amplitude. These settings do not correlate to any specific amplitude and may
vary between instruments.
Shape
square wave
Frequency
153 Hz to 5 MHz
Amplitude
5
positions
Summary of Contents for 71501D
Page 1: ...Agilent 71501D Eye Diagram Analysis User s Guide...
Page 7: ...1 Getting Started...
Page 34: ...Getting Started Steps for Setting Up Eye Diagram Analysis...
Page 35: ...2 Application Tutorials...
Page 43: ...2 9 Application Tutorials Tutorial 2 Measure in Optical Power Units...
Page 48: ...Application Tutorials Tutorial 4 Measure Laser Turn on Delay Laser Overshoot and Turn On Delay...
Page 80: ...Application Tutorials Tutorial 11 Create a Horizontal Histogram...
Page 81: ...3 Eye Diagram Analyzer Reference...
Page 113: ...3 33 Eye Diagram Analyzer Reference Applying Mask Testing The Mask Setup Submenu...
Page 151: ...4 Programming Commands...
Page 163: ...5 Specifications and Characteristics...
Page 170: ...Specifications and Characteristics Declaration of Conformity Declaration of Conformity...