
5
Operating in EI Mode
Set the IP Address and View TQ Parameters with the WUI
7000/7010 Series TQ GC/MS Operating Manual
97
Set the IP Address and View TQ Parameters
with the WUI
The Web User Interface (WUI) is used by Agilent field support for setup and
monitoring of the instrument. There are a couple of things available in this WUI that
are useful and sometimes essential for operation. These include setting the TQ IP
address and monitoring instrument parameters during pumpdown without the
Agilent MassHunter GC/MS Acquisition software use. These procedures assume
that the operator has access to a PC located on the same LAN subnet as the TQ.
To change the network settings of the TQ
The GC firmware does not support LVDS communications with the 7000/7010
series TQ MS at this time so you must use the Web User Interface (WUI) to
configure the TQ network settings.
1
Press the
MS On/Off
switch to start the instrument.
2
Enter the instrument IP address into a PC web browser URL to display the
WUI Main Menu page. (See
If the IP address was never configured, the factory defaults for the IP
address, gateway, and subnet are 192.168.254.12, 0.0.0.0, and
255.255.255.0 respectively.
Figure 26.
WUI main menu