7
General Maintenance
To Disassemble the EI HES
166
7000/7010 Series TQ GC/MS Operating Manual
11
Remove the five lenses from the lens insulator/holder.
CAUTION
Use care when removing the lenses from the lens insulator casing. Putting
excessive stress on this casing can break or crack it. If this happens, do not
attempt to operate with a defective lens insulator, it must be replaced.
Figure 52.
Exploded parts view of the EI HES
1
2
8
16
17
15
9
4
3
11
10
6
5
7
13
12
14
8
12
Table 15 Parts list for EI HES (
Item number
Item description
Part number
1
Source finger grip
G7002-20008
2
Filament block
G7002-20019
3
Extractor lens (5)*, with 3 mm opening
G7004-20061
4
Ceramic insulator for extractor
G7002-20064