Operating in EI Mode
To Pumpdown the MS
122
7000/7010 Series TQ GC/MS Operating Manual
11
Select the
Manual Tune
tab.
12
Click
Pumpdown
. You are prompted if you want to load high temperature
settings for the ion source and MS1 and MS2 quads.
13
Click
Yes
to set the source/MS1/MS2 at 230/150/150 or click
No
to set them
at 100/100/100.
14
The
Pumpdown
dialog displays. (See
.) You are prompted to make
sure the vent valve is closed.
15
Click
OK
to close the Pumpdown manual step dialog and start the
Pumpdown.
Figure 41.
Pumpdown status