250
Volatiles Interface
Theory of operation
2 of 30
Jun 2001
Agilent 6890 Gas Chromatograph Service Manual
Figure 250-1
VI flow control diagram: split mode splitless mode (Idle or after
sampling end)
Splitless mode
When you introduce a sample, the solenoid valve remains closed while the
sample enters the interface and is transferred to the column. At a specified
time after the sample is introduced, the solenoid valve opens.
Before Pre Run
, when the GC is preparing for sample introduction, total flow
to the interface is measured by a flow sensor and controlled by a proportional
valve. Column flow is controlled via back-pressure regulation. See
Sampler
Trap
Valve
open
PV2
SPR
SPR
Vent
Split
Vent
FS
PV1
Total flow
control loop
PS
To Column
Supply